NT2A Vital Signs Monitor
CE
|
NT2A:
Vital Signs Monitor
- Bright, Easy-to-Read Large LED Display
- Durable, Compact and Lightweight
- Long-life Internal Rechargeable Battery for
Uninterrupted Monitoring
- Accurate and Reliable During Low Perfusion and
Montion
- Audible/Visual Three Level Alarms with Alarm Volume
Control
- 72-hour Data Memory Allowing
On-screen Viewing of SpO2
and Pulse Rate Trends
- Suitable for Adult, Pediatric and Neonate
- SpO2
Accuracy:
No Motion:
Adults, Pediatrics: 70¡«100% ¡À2%
0¡«69%
unspecified
Neonates:
70¡«100% ¡À3%
0¡«69%
unspecified
Motion:
Adults, Pediatrics: 70¡«100% ¡À3%
0¡«69%
unspecified
Neonates: 70¡«100%
¡À4%
0¡«69%
unspecified
Low Perfusion:
Adults, Pediatrics: 70¡«100% ¡À3%
0¡«69%
unspecified
Neonates: 70¡«100%
¡À4%
0¡«69%
unspecified
- Pulse Rate Accuracy
No Motion:
Adults, Pediatrics, Neonates: ¡À3%
Motion:
Adults, Pediatrics, Neonates: ¡À5%
Low Perfusion:
Adults, Pediatrics, Neonates: ¡À5%
- Displays and Indicators
SpO2, Pulse Rate and
Pulse Strength Bar
Pulse Search Indicator
Low Perfusion and Motion Warning Indicator
Alarm Silence Indicator
AC Power and Battery Charge Indicator
Low Battery Warning Indicator
Sensor Disconnect and Sensor off Warning Indicator
- Alarm
Audible/Visual Alarms for High/Low SpO2 and PR
Audible/Visual Alarms for Low-Battery
Alarm Limits Adjustable
Alarm Volume Adjustable
Pulse Search Indicator
Alarm Silence Interval:30,60,90and120sec and OFF
Low Perfusion and Motion Warning Indicator
Sensor Disconnect Alarm
Sensor off Alarm
- Performance Measurement
Range
SpO2:0¡«100%
Pulse Rate: 20bpm¡«250bpm
- Environment
Opreation Temperature: 5¡«40¡æ
41~104¨H
Humidity: ¡Ü80%
Altitude: -390¡«5,000m(-1,280¡«16,404ft)
Transport/Storage Temperature:
-20¡«+70¡æ (-4¨H¡«158¨H)
Humidity: ¡Ü95%
- Output
RS-232 Serial Port for Digital Output
Nurse Call Function
- Electrical
Instrument
Power Requirements:
100~240VAC, 50/60Hz, 20VA
Fuse Rating: 1.0A
- Battery
Internal Battery: 12V 3.6Ah¡¡
Battery Capacity: ¡Ý 24 hours
- Physical Characteristics
Dimensions:
264mm(W) x 95mm(H) x 172mm(D)
[10.39in(W) x 3.74in(H) x 6.77in(D)
Net Weight: 2.5kg/5.5ib
Gross Weight: 3.5kg/7.7ib
- Compliance
SpO2: ISO9919:2005¡¡
Safety Standards:
En60601-1: 1990+A1: 1993+A2:
1995+A13: 1996 Type BF-Applied
IPX1 Approved
Alarm: IEC60601-1-1-8: 2005
EMC:
EN60601-1-2: 2001, Group1 ClassA
Environment: WEEE(2002/96/EC)
- Thermal Arrary
Recorder(Optional)
Prints Graphical and Tabular Trend
|
|
|