Feeding Pump
Function and Features The pump can be used for continuous feeding.
Rate: 1-400ml/h.
All kinds of alarm functions.
IV set: Universal. All brands of national standard are compatible.
Bolus: 400ml/h.
Technical parameters of Enteral Feeding Pump
|
|
Bolus rate |
400ml/h |
Maximum size of outer shell |
120*140*190mm (length*width*height) |
Maximum weight |
<2.5kg |
Infusion rate |
1-400ml/h |
Volume limit |
1-9999ml |
Accuracy |
+10% |
Power supply |
AC100~220V, 50/60Hz |
Displayed information |
Flow rate, volume limit, accumulated volume, power supply indicator, bed No. |
Battery |
Rechargeable lithium polymer battery, 7.4V, 1600mAh |
Alarm functions |
Over, low battery, control abnormity, no AC power supply |
Battery recharge |
When the pump is connected to the AC power supply, turn on the power and the battery will automatically recharge (about 8~14 hours to recharge fully.) |
Fuse |
T2AL250V~ |
Maximum power consumption |
25W, can run for 4h after being recharged. |
Classification |
II Class-BF |
Storage conditions |
Environment temperature -40¡ãC~55¡ãC, air pressure 50kPa ~106kPa, relative humidity ¡Ü95% |
Outer shell material |
ABS plastic |
Operating conditions |
Environment temperature 5¡ãC ~ 40¡ãC, air pressure 86kPa ~106kPa, relative humidity ¡Ü80% |
Applicable infusion sets |
All standard infusion sets |
Applied standards |
IEC 60601-1: 2005
IEC 60601-2-24: 1998
IEC 60601-1-2: 2006
ISO780-1997
ISO15223: 2000
ISO8536-4: 2004
ISO/IEC: Guide No. 37¡¶Consumer Goods Usage Instructions¡·
|
|